Internal Quality Control (IQC): Principles, Levey–Jennings Charts and Troubleshooting

Purpose and learning objectives

Internal quality control (IQC) monitors analytical performance using materials measured within the laboratory. It helps staff recognise departures from an established process and decide when investigation is required. After studying this guide, you should be able to read a control chart, distinguish selected error patterns and describe a procedure-led response.

Calibration is different from IQC

A calibrator establishes the relationship between signal and assigned quantity. A control material provides continuing evidence about measurement performance; it should be used according to the laboratory’s validated plan. Passing calibration does not mean later measurements remain stable, and acceptable IQC cannot exclude all patient-specific interference (WHO, 2011).

Control materials

Use materials appropriate for the assay, matrix and clinically relevant concentrations. Preparation, stability, storage and lot identification require control. Assigned values and local verification must be appropriate to the method. A new lot may require overlap studies and review of target values; do not simply move the target to accommodate unexplained drift.

Mean, standard deviation and coefficient of variation

The mean describes the centre of repeated values. Standard deviation (SD) describes dispersion in the original measurement units. For a sample of observations, the usual sample SD uses n − 1 in the variance denominator. Laboratory target means and SDs should come from an appropriate, documented establishment process.

CV (%) = (SD ÷ mean) × 100

For an established mean of 6.0 mmol/L and SD of 0.2 mmol/L, CV = 3.33%. CV expresses variation relative to the mean and is unstable when the mean is near zero. It does not measure bias against a reference value (NIST, n.d.).

A Levey–Jennings chart

The horizontal axis shows successive control measurements and the vertical axis shows their values. Reference lines mark the target mean and ±1, ±2 and ±3 SD. These are analytical control limits, not a patient reference interval.

Fictional glucose IQC: target mean 6.0 mmol/L and SD 0.2 mmol/L. Ten sequential results are plotted with lines from minus 3 SD to plus 3 SD. Run 10 is 6.8 mmol/L, or plus 4 SD.
Fictional glucose IQC: target mean 6.0 mmol/L and SD 0.2 mmol/L. Ten sequential results are plotted with lines from minus 3 SD to plus 3 SD. Run 10 is 6.8 mmol/L, or plus 4 SD.

Accessible dataset: runs 1–10 are 6.0, 6.1, 5.9, 6.0, 6.2, 6.1, 6.0, 6.3, 6.4 and 6.8 mmol/L. The established target is 6.0 mmol/L; SD is 0.2 mmol/L. Lines are 5.4 (−3 SD), 5.6 (−2 SD), 5.8 (−1 SD), 6.0 (mean), 6.2 (+1 SD), 6.4 (+2 SD) and 6.6 (+3 SD). These limits are stipulated for teaching, not estimated from this short dataset.

Run 10 is 4 SD above the target: (6.8 − 6.0)/0.2 = 4. In the accompanying fictional activity, this exceeds the laboratory’s applicable acceptance criteria. Interpretation of the whole sequence still requires the defined QC strategy and information from other levels and assays.

Random and systematic error

Random error increases unpredictable scatter. A systematic change may appear as a shift in the centre or a progressive trend. Patterns suggest possible mechanisms but do not confirm a cause. For example, variability may prompt assessment of aspiration or mixing, while a shift after a reagent change prompts review of lot verification and calibration.

Common control rules and their limitations

Examples include 1₃s (a result beyond 3 SD), 2₂s (two relevant results beyond 2 SD on the same side) and R₄s (a difference exceeding 4 SD between relevant results within a run). In the original multirule approach, 1₂s is a warning that prompts further review. The number of controls, definition of a run and combination of rules matter (Westgard, n.d.).

There is no universal mandatory Westgard combination for every assay. Rules balance error detection against false rejection. The laboratory selects and validates its strategy using assay performance, clinical risk and applicable requirements. Do not apply within-run rules indiscriminately across unrelated days or concentration levels.

Responding to unacceptable IQC

  1. Follow the applicable procedure for holding affected results and notifying the responsible staff.
  2. Check control preparation, lot, expiry, storage, target selection and transcription.
  3. Review reagent lots, calibration, maintenance, alarms and the pattern across other controls and assays.
  4. Investigate the evidence and document corrective action. Do not rerun repeatedly until a control happens to pass.
  5. Complete the defined checks for returning the assay to service. Assess the time window and possible impact on previously measured specimens.

Documentation and escalation

Record the observation, affected assay and period, investigation, confirmed cause if known, action, verification and authorisation. Escalate persistent or unexplained failures through the local quality system. Corrected reporting and clinical communication require the appropriate authorised process.

Apply these ideas in the fictional reflective CPD activity.

IQC, EQA and accreditation

IQC supports frequent monitoring within the laboratory. External quality assessment (EQA) provides comparison using an external programme, with interpretation against appropriate assigned values or peer groups. They offer complementary evidence; neither replaces the other.

UKAS accredits medical laboratories against ISO 15189 requirements for quality and competence. ISO 15189:2022 is the current edition discussed here. Accreditation relates to the laboratory’s defined scope, rather than approval of every individual educational resource or numerical result. Consult current UKAS publications and your laboratory quality system (UKAS, n.d.).

Check your understanding

1. Is a calibration curve an IQC chart?

No. Calibration establishes a measurement relationship. An IQC chart tracks control performance over time.

2. What is the CV for a mean of 6.0 mmol/L and SD of 0.2 mmol/L?

(0.2/6.0) × 100 = 3.33%. This describes relative dispersion, not accuracy by itself.

3. Does a +4 SD result prove the reagent lot is defective?

No. It demonstrates a departure from the stipulated target and prompts action under the QC plan. Investigation is required to determine the cause.

References and further reading