Learning objectives
- Distinguish absorbance from transmittance.
- Apply A = εlc with consistent units and appropriate assumptions.
- Explain the roles of the optical system, blanks, calibration and IQC.
- Recognise interference and connect a measurement principle with an assay-specific method.
What is spectrophotometry?
Spectrophotometry measures the interaction between light and matter at selected wavelengths. In solution absorbance measurements, the instrument compares light transmitted through the sample with an appropriate reference. Molecules absorb particular wavelengths according to their electronic structure. The recorded spectrum can therefore support identification and quantification (IUPAC, n.d.).
Light can also be reflected or scattered. A detector cannot assume that every reduction in transmitted light arises from the intended reaction product. The sample, vessel and optical conditions must be controlled.
Absorbance and transmittance
Transmittance is the fraction of incident light transmitted: T = I/I₀. Percentage transmittance is 100 × T. Absorbance is A = −log₁₀(T) = log₁₀(I₀/I). It is dimensionless. If T = 0.10, 10% of the reference light is transmitted and A = 1.00. A doubling of absorbance does not mean a doubling of percentage absorption (Shimadzu, n.d.a).
The Beer–Lambert law
A = εlc
- A: absorbance, dimensionless.
- ε: molar absorption coefficient at a specified wavelength, commonly L mol⁻¹ cm⁻¹.
- l: optical path length through the absorbing medium, in cm for these units.
- c: amount concentration of the absorbing species, in mol/L.
With suitable monochromatic radiation, a homogeneous sample and a stable chemical species, absorbance is proportional to path length and concentration. The coefficient depends on the species, wavelength and conditions; it is not a universal constant for all assays. In SI units, ε may be expressed in m²/mol with l in metres and c in mol/m³ (IUPAC, n.d.).
Worked example: for a fictional absorbing species with ε = 10,000 L mol⁻¹ cm⁻¹, l = 1 cm and c = 0.00005 mol/L, A = 10,000 × 1 × 0.00005 = 0.50. Conversely c = A/(εl). This example illustrates units, rather than a particular patient assay.
Components of a spectrophotometer
The optical system includes a light source, a wavelength-selection system, a sample compartment or reaction cuvette, and a detector. Electronics convert the detector signal into a reported measurement. Depending on the design, wavelength selection may use filters or a monochromator, and a reference beam may be measured separately. The cuvette material must transmit the chosen wavelength (Shimadzu, n.d.b).
Blanks, calibrators and quality control
A blank corrects for specified background contributions from the vessel, solvent or reagents. It is selected by the method and does not remove every possible patient-specific interference. A calibrator establishes the relationship between the analytical signal and the assigned quantity. IQC monitors the performance of the configured measurement process. Acceptable calibration does not replace ongoing IQC (WHO, 2011).
Applications in automated Clinical Chemistry
An automated analyser dispenses sample and reagent into a reaction vessel, controls timing and temperature, and reads optical signals. In an endpoint method, a defined measurement is taken after an agreed reaction interval. A kinetic method evaluates signal change over time. Concentration or enzyme activity is derived using the method’s calibration or validated factor.
Assay-specific example: the Siemens ADVIA ALT method couples the ALT reaction to lactate dehydrogenase. NADH is consumed as pyruvate is converted to lactate; the decline in absorbance at 340/410 nm relates to ALT catalytic activity. This is a coupled enzyme measurement, not direct absorption by ALT protein. Wavelengths, reagent composition and pyridoxal phosphate activation vary by assay (Siemens Healthineers, 2020).
Other photometric approaches include colour-forming reactions and UV coenzyme measurements. A sample’s reported concentration need not equal the concentration of the absorbing reaction product without accounting for reaction stoichiometry and dilution. Turbidimetric measurements rely on scattering and require their own validated calibration model.
Follow the analyser sample-to-result pathway or read the existing ALT test guide.
Interference and non-linearity
Haemoglobin, bilirubin and lipaemic turbidity can affect optical assays, but the direction and magnitude are method dependent. Consult the applicable instructions for use and the laboratory’s interference policy. A haemolysis index is evidence for assessment, rather than a universal instruction to reject every affected sample.
Stray light can cause measured absorbance to underestimate the true value, particularly at high absorbance. Finite spectral bandwidth, chemical changes at high concentrations and detector limits can also disturb linearity. Use the validated analytical measuring range and approved dilution protocol; do not extrapolate a calibration curve beyond its verified limits (Shimadzu, n.d.c).
Practical troubleshooting
- Review error flags, reaction curves and the pattern across samples and QC levels.
- Check sample and reagent volumes, bubbles, cuvette condition and optical cleanliness.
- Review reagent lot, expiry, storage, calibration and maintenance records.
- Assess specimen quality and assay-specific interference information.
- Follow the local procedure for corrective action, authorisation, repeat testing and documentation. Do not repeatedly rerun a control simply to obtain an acceptable value.
Check your understanding
1. What is the absorbance when 10% of the reference light is transmitted?
T = 0.10, so A = −log₁₀(0.10) = 1.00.
2. What happens if the optical path length doubles under suitable Beer–Lambert conditions?
Absorbance doubles at the same concentration and coefficient. Confirm that the measurement remains within the validated range.
3. Why does acceptable calibration not replace IQC?
Calibration sets the measurement relationship at a point in time. IQC provides continuing evidence about analytical stability and may reveal changes after calibration.
References and further reading
- IUPAC (n.d.) Beer–Lambert law. Gold Book. doi:10.1351/goldbook.B00626.
- Shimadzu (n.d.a) How does a spectrophotometer measure a sample?
- Shimadzu (n.d.b) The structure of a spectrophotometer.
- Shimadzu (n.d.c) How does stray light influence high absorbance measurements?
- Siemens Healthineers (2020) Alanine Aminotransferase (ALT): ADVIA 1800/2400 Chemistry Systems. Verify the current local assay instructions before practice.
- WHO (2011) Laboratory quality management system: handbook.
Online sources accessed 21 November 2026. Diagram is an original LabPathPro educational illustration.